• Room209-210,Lyshine Plaza,Shijingshan Zhongguagncun Technology Park
  • beijing, Beijing, 100043
  • Kinija
  • Telefonas:+861053367875
  • Faksas: ---.---.-----
  • URL:

ANESTHESIA MACHINE MODEL:DA2000

Kvėpavimo įranga

PAGAL UŽKLAUSIMĄ

Naujas

Nėra

Nėra

Teirautis

Produkto aprašymas

Medical Anesthesia Machine, Open Anesthesia Machine, Touch Screen Anesthesia Machine Supplier

Feature:

More convenient operating and display systems for clinical operations,suitable for a wide range of anesthesia at adult,pediatric

Comprehensive Safety&Alarm system,make sure the operation accuracy and stability

8 inch TFT display with touch screen,shuttle knob design, quick and efficient operation

Integrated absorber design with good cable management solution,convenient to discharge and install timely..

Powerful ventilation mode meet all complex clinical requirement

Three levels alarm system, visual and sound alarm information.

The core components are imported from top international and original, and the industrial control chip is chosen

High precious sensor and valve technology make sure the pressure ,flow and volume monitor and control at high safety, stability and precision

High precious anesthesia vaporizer with temperature,flow and pressure compensation

Top-light system design provide more good vision

Built-in backup battery provide the emergency power supply to the unit.

Low O2 pressure alarm and N2O cut-off protection.

Technical Specifications

Display:8 " TFT, with touch screen.

Working Mode:Closed,Semi-closed,,semi-open system,Open

Ventilation Mode:VCV,PCV,SIMV-VC,SIMV-PC,PSV,Manual/SPONT.

Optional: SIMV-PRVC, PRVC

Waveform:Pressure,Flow-rate,Volume

Loop:PV loop, PF loop,FV loop

Safety Pressure:Does not exceed 6 KPa.

Battery Type:Li-ion,work time more than 2 hours

Parameter Setting Range:

Tidal Volume:Range: 20 - 1500ml;

Respiratory rate::1~100 bpm;

Inspiratory time:0.1 ~ 10.0 s;

Respiratory ratio: 9.9:1 to 1:9.9;

Percentage of inspiratory pause:0 to 50%;

PEEP:OFF, 3 ~ 20 cmH2O

Pressure Support:0 ~ 60 cmH2O

Pressure Control:5 ~ 60 cmH2O

Flow trigger: 0.5 ~ 20L/min

Pressure Trigger: -1 ~ -20 cmH2O

PSV Insp Termination Level:25%

Monitoring Parameter

Inspiratory tidal volume,Expiratory tidal volume,Minute ventilation,Spontaneous minute ventilation,Respiratory rate,Spontaneous breathing frequency,Respiratory ratio,Peak airway pressure,Mean airway pressure,PEEP,Inspiratory plateau pressure,FiO2(optional)

Alarm Setting:

Tidal volume:High 0 ~ 2000 mL, OFF;Low 0 ~ 1500 mL

Minute ventilation:High 1 ~ 99 L;Low 0 ~ 98 L

Respiratory rate:High 1 ~ 100bpm;Low 0 ~ 99 bpm

FiO2(optional):High 19 ~ 100%;Low 18 ~ 99%

Airway pressure:High 6 ~ 99 cmH2O, ;Low 0 ~ 9 8 cmH2O

FiCO2 (optional):High 0.1 to 13.3%; Low 0 to 13.2%

Continuous Positive Airway Pressure:Greater than (PEEP +15) cmH2O,

15 s sustained when an alarm is triggered.

Negative pressure alarm:Airway pressure is less than (-10) cmH 2 O.

Apnea alarm:Setting time is 10 ~ 60 s

Gas deficiency alarm:less than 280 kPa.

AC power failure alarm:The mains breaks down or the power cord disconnects

Low battery alarm: Time is 20 min.

Alarm battery is exhausted: time is 10 min.

Silent Alarm:≤ 120 s

O2 sensor failure:FIO2