ANESTHESIA MACHINE MODEL:DA2000
Beskrywing van die produk
Medical Anesthesia Machine, Open Anesthesia Machine, Touch Screen Anesthesia Machine Supplier
Feature:
More convenient operating and display systems for clinical operations,suitable for a wide range of anesthesia at adult,pediatric
Comprehensive Safety&Alarm system,make sure the operation accuracy and stability
8 inch TFT display with touch screen,shuttle knob design, quick and efficient operation
Integrated absorber design with good cable management solution,convenient to discharge and install timely..
Powerful ventilation mode meet all complex clinical requirement
Three levels alarm system, visual and sound alarm information.
The core components are imported from top international and original, and the industrial control chip is chosen
High precious sensor and valve technology make sure the pressure ,flow and volume monitor and control at high safety, stability and precision
High precious anesthesia vaporizer with temperature,flow and pressure compensation
Top-light system design provide more good vision
Built-in backup battery provide the emergency power supply to the unit.
Low O2 pressure alarm and N2O cut-off protection.
Technical Specifications
Display:8 " TFT, with touch screen.
Working Mode:Closed,Semi-closed,,semi-open system,Open
Ventilation Mode:VCV,PCV,SIMV-VC,SIMV-PC,PSV,Manual/SPONT.
Optional: SIMV-PRVC, PRVC
Waveform:Pressure,Flow-rate,Volume
Loop:PV loop, PF loop,FV loop
Safety Pressure:Does not exceed 6 KPa.
Battery Type:Li-ion,work time more than 2 hours
Parameter Setting Range:
Tidal Volume:Range: 20 - 1500ml;
Respiratory rate::1~100 bpm;
Inspiratory time:0.1 ~ 10.0 s;
Respiratory ratio: 9.9:1 to 1:9.9;
Percentage of inspiratory pause:0 to 50%;
PEEP:OFF, 3 ~ 20 cmH2O
Pressure Support:0 ~ 60 cmH2O
Pressure Control:5 ~ 60 cmH2O
Flow trigger: 0.5 ~ 20L/min
Pressure Trigger: -1 ~ -20 cmH2O
PSV Insp Termination Level:25%
Monitoring Parameter
Inspiratory tidal volume,Expiratory tidal volume,Minute ventilation,Spontaneous minute ventilation,Respiratory rate,Spontaneous breathing frequency,Respiratory ratio,Peak airway pressure,Mean airway pressure,PEEP,Inspiratory plateau pressure,FiO2(optional)
Alarm Setting:
Tidal volume:High 0 ~ 2000 mL, OFF;Low 0 ~ 1500 mL
Minute ventilation:High 1 ~ 99 L;Low 0 ~ 98 L
Respiratory rate:High 1 ~ 100bpm;Low 0 ~ 99 bpm
FiO2(optional):High 19 ~ 100%;Low 18 ~ 99%
Airway pressure:High 6 ~ 99 cmH2O, ;Low 0 ~ 9 8 cmH2O
FiCO2 (optional):High 0.1 to 13.3%; Low 0 to 13.2%
Continuous Positive Airway Pressure:Greater than (PEEP +15) cmH2O,
15 s sustained when an alarm is triggered.
Negative pressure alarm:Airway pressure is less than (-10) cmH 2 O.
Apnea alarm:Setting time is 10 ~ 60 s
Gas deficiency alarm:less than 280 kPa.
AC power failure alarm:The mains breaks down or the power cord disconnects
Low battery alarm: Time is 20 min.
Alarm battery is exhausted: time is 10 min.
Silent Alarm:≤ 120 s
O2 sensor failure:FIO2