Thermal Shock Chamber

Dzīvnieku farmācijas

PĒC PIEPRASĪJUMA

Jauns

Nav pieejams

Nav pieejams

Jautāt

Apraksts

Thermal Shock Chamber

LIB thermal shock chamber for sale has types: basket type, three-room type, and horizontal movement type. Including air to air hot cold shock test, air to air thermal shock test and liquid to liquid thermal shock test. According to different test requirements, customers choose suitable thermal shock test chambers for testing.

Model: TS-162, TS-340, TS-500, TS-1000

Standard: AAMA501, ASTM D4169-16, IEC 61215, IEC 61646, IEC61108, IEC62688, MIL-STD-813D

Thermal Shock Chamber Features and Benefits

1. The thermal shock chamber has upper and lower basket type and static damper type.

2. The high temperature chamber can realize air-to-air thermal shock testing, air to air thermal shock testing and liquid to liquid thermal shock testing.

3. From obtaining customer needs, to discussing solutions, production, transportation, installation and commissioning, LIB can realize one-stop solutions.

4. The thermal shock environmental chambers, a kind of portable temperature humidity chamber, are mainly used in high and low temperature shock test of automobiles, EV batteries, electronic products and etc.

Specifications of Thermal Shock Chamber

Model TS-162 TS-340 TS-500 TS-1000

Internal Dimensions (mm) 300*300*250 450*450*360 650*650*500 850*850*700

Overall Dimension (mm) 1560*870*1545 1710*1020*1845 1910*1220*2265 2110*1420*2665

Interior Volume (mm) 22L 72L 211L 505L

Loading Capacity 20kg 30kg 50kg 60kg

Pre-heat Room Upper limit Temperature +220℃

Heating Time Ambient ~ + 200℃, within 30 minutes

Pre-cool Room Lower limit Temperature -75℃

Cooling time Ambient ~ -70℃, within 30 minutes

Test Room High Temperature Exposure Range Ambient +20 ~ +200°C

Low temperature Exposure Range –65 ~ -5°C

Temperature Fluctuation ≤±0.5℃

Temperature Deviation ≤±3 ℃

Temperature Recovery Time Within 5 minutes

Cooling System Mechanical compression refrigeration system

Controller Programmable color LCD touch screen controller, Ethernet connection

Safety Device Over-temperature protection, Over-current protection; Refrigerant high-pressure protection; Earth leakage protection

Exterior Material Steel Plate with protective coating

Interior Material SUS304 stainless steel

Observation Window Interior lighting, double-layer thermo stability silicone rubber sealing

Standard Configuration 2 shelves

Thermal Shock Chamber Construction

LIB 2-zone hot cold thermal shock test chamber is available in small capacity and large capacity to meet different testing requirements. Test can be performed during thermal shock testing from -70 to +200°C.The specimen automatic transferred from cold chamber to hot chamber by basket. The basket slides vertically and smoothly through rails, to make the specimen is exposed to the two chambers.

The construction of a temperature shock test chamber typically includes the following components:

1. Insulated chamber: The main body of the chamber is typically constructed from stainless steel or other high-quality materials to provide insulation and durability. The chamber is divided into two compartments – a hot compartment and a cold compartment.

2. Heating and cooling systems: The hot compartment is equipped with a heating system, such as electric heaters or gas burners, to rapidly raise the temperature. The cold compartment is equipped with a cooling system, such as compressors or liquid nitrogen, to rapidly lower the temperature.

3. Temperature control system: A PID temperature control system is used to precisely regulate the temperature in each compartment. This system may include sensors, heaters, and coolers to maintain the desired temperature levels.

4. Transfer mechanism: A transfer mechanism, such as a conveyor belt or robotic arm, is used to move the test samples between the hot and cold compartments. This allows for rapid temperature changes and ensures consistent testing conditions.