EST-HS2 Manual Durometer Test Stand
Produktbeskrivelse
EST-HS2 durometer operating stand is a test stand for durometer. The durometer stand is a workbench for loading Shore hardness tester testing. Using a three-point leveling stainless steel work platform, this durometer test stand is more stable and its measurement result is more accurate. The overall structure is reasonable.
Compared with shore hardness gauge, EST-HS2 durometer operating test stand can test various types of shore hardness.
EST-HS2 Manual Durometer Test Stand Features
The durometer test stand can test various types of shore hardness.
Cam-type loading and unloading make the test process more stable.
Operation type: the indenter pressed against the sample
The measuring rod of the durometer stand can be adjusted by lifting.
Workbench adjustment: three-point leveling.
Durometer Operating Stand Specification
Operating Type Inverter to specimen
Loading Method Cam loading
Loading Stroke 18 mm
Specimen Thickness 60 mm
Throat Depth 65 mm
Support Table(dia.) Φ116 mm
Lifting Adjustment Screw adjustment
Table Adjustment Three-point leveling
Net Weight 5.6 kg (excluding S weight)
Package Weight 7.6 kg (excluding D weight)
Durometer Operating Test Stand Operation
1. Unpacking
1.1 Take out the durometer test stand from the carton and remove the two packing foam inserts
1.2 Unbundle the webbing strap and remove the foam sheet and foam pad
1.3 Install the handle of durometer stand
2. Weights Selection of Durometer Stand
3. Simple Adjust for Support Table
In most applications, simple adjustments are usually sufficient for the durometer operating stand.
3.1 Release Tighten screw, turn height adjust knob, let durometer stand move down to contact on the support table.
3.2 Adjust 2 Vernier Knobs. Let the face of the support table be fully fitted to the bottom of the durometer operating stand.
4. Install Durometer Operating Test Stand
4.1 Lift Handle to put down durometer stand
4.2 Tighten Durometer Stand through Tighten Knob
5. Fine Adjust for Support Table
When accurate measurements or calibration are performed, fine adjustments to the support table may be required.
5.1 Set the test mode of the durometer test stand to Timing Mode and set its Dwell Time as long as possible, such as 99 s. (Lengthening the time period for reading lock)
5.2 Put standard thickness block between press foot and the support table.
5.3 Lift the handle to make the indenter of the durometer stand press onto the support able through the hole on the block.
5.4 Now the readings are displayed on the durometer test stand. Adjust the Vernier Knobs to make the reading as larger as possible to near 50HS.
Note:
1. If adjustment is not finished in setting time, lift the durometer operating stand then press again for renewing the timing.
2. Because the indenter is often pressed into the table, the reading does not must be "50.0 HS", It is usually less than 50.0 HS. Just adjusting the table to the location of the maximum reading.
3. Too large or too fast adjusting may damage the indenter (especially for type D). So it is better to pre-adjust with the durometer test stand loaded. Or lift up and press down the durometer stand in adjustment.
As a measuring instruments manufacturing company, we have types of force gauge meter for sale, if you have needs, please contact us.