XKDL-8000 Automatic Infrared Measuring Sulfur Analyzer
Ürün tanımı
XKDL-8000 Automatic Infrared Measuring Sulfur Analyzer
Sulphur Analyzer
Product Name: Automatic infrared measuring sulfur analyzer
Product Type: XKDL-8000
Scope:
Coal, electricity, chemical, geological surveying, environmental protection, quarantine inspection, research institutions and other related industries and sectors. Provide accurate data for total sulfur arbitration.
Features:
1. The furnace insulation material: A grade mullite, thermostats with a long and stable with double high temperature combustion tube for secondary high-temperature combustion of the sample.
2. Connection mode: VB-CAN card, use shielding and lightning protection circuit, anti-magnetic interference devices, anti-stimulus interference.
3. Temperature control mode: PID algorithm, anti-overshoot heating, safe, precise temperature control, heating speed.
4. Gas components: the use of double flow control technology enables a stable supply of oxygen flow, high and low pressure alarms. And unique purge injector device and an air curtain, so that the outer isolation.
5. Using self-developed ultra-low-drift infrared pond, a working day only once drift correction to meet the requirements, the operator switched to a day without interruption to carry out sample testing once corrected, without frequent calibration during the experiment that isYou can have accurate test results, high test efficiency.
6. Hardware bit machine effectively controlled by the host computer unilateral action does not cause uncontrollable. Fault alarm to protect themselves. Independent development of dust, in addition to the water treatment system.
7. Accurate and reliable data, the system automatically uses a unique system of data analysis standard linear error correction.
Technical parameters:
1. Test method: IR method.
2. Sulfur measurement range: (0.001~99.9)%, sulfur Resolution: 0.001%
3. Test time: (120) S / sample, sample weight: 100-500mg
4. The heating rate: 50-100 ℃ / min in accordance with the relevant national standards
5. Precision: within the range of uncertainty in the standard sample
6. Operating furnace temperature: 1350 ± 5 ℃, control accuracy: ± 1 ℃
7. Voltage: 220 ± 22VAC 50 ± 1HZ power: